10 kg Active Hydrogen Maser

Model: ECNU-2310max

The JLT-2310 MAX (ECNU-2310 MAX) active hydrogen maser is a 12 kg enhanced high-performance ultra-compact active hydrogen maser offering excellent frequency stability. It uses the most advanced process integration technology in a standard 4U rack-mount chassis. Its specifications exceed those of the 1008M, and it is a 100% domestically produced substitute. The JLT-2310 MAX is well suited to automated time service and test systems. It provides RS-232C, USB and LAN interfaces for remote access to operating status parameters and control.

Applications

  • · National time service;
  • · Space tracking and navigation;
  • · Metrology and testing;
  • · Scientific research.

Specifications

Waveform Qty Amplitude Bandwidth, ps Rise time, ns Interface / Load
5 MHz (sine) 2 1±0.2VrMs - - SMA/50 Ω
10 MHz (sine) 2 1±0.2VrMs - - SMA/50 Ω
Performance
Relative frequency accuracy ± 3×10⁻¹³ ≤ ±1×10⁻¹³
Frequency stability (Allan deviation at 25±1 °C, environmental effects excluded) 1 s ≤ 3.0×10⁻¹³
10 s ≤ 5.0×10⁻¹⁴
100 s ≤ 2.0×10⁻¹⁴
3600 s ≤ 3.0×10⁻¹⁵
86400 s ≤ 8.0×10⁻¹⁶
Phase noise
(dBc/Hz)
Offset @5MHz @10MHz @100MHz
1 Hz ≤ -105 ≤ -100 ≤ -80
10 Hz ≤ -130 ≤ -120 ≤ -100
100 Hz ≤ -145 ≤ -140 ≤ -120
1 kHz ≤ -155 ≤ -145 ≤ -150
10 kHz ≤ -155 ≤ -145 ≤ -150
Frequency drift (after 1 year of operation) ≤ 1×10⁻¹⁵/day
Magnetic sensitivity ≤ 1×10⁻¹⁴/Gauss
Temperature sensitivity ≤ 1×10⁻¹⁴/℃
Physical Specifications
Dimensions (W × H × D) 470 mm × 200 mm × 530 mm
Weight ~12kg
Operating temperature range (±0.5 °C/hour) 10–35 °C (recommended +22 °C)
Relative humidity < 80% at +25 °C