CN|EN
> Products > Product Detail

10 kg Active Hydrogen Maser

Model: ECNU-2310Pro

The JLT-2310 PRO (ECNU-2310 PRO) active hydrogen maser is an ultra-compact active hydrogen maser offering excellent frequency stability. It uses the most advanced process integration technology in a standard 4U rack-mount chassis. Its specifications exceed those of the VCH1008, and it is a 100% domestically produced substitute. The JLT-2310 PRO is well suited to forming part of an autonomous cooperative swarm timing system and integrates readily with various platform systems. RS-232C, USB and Ethernet (LAN) interfaces can be supplied to order for accessing operating status and control parameters.

Applications

  • · National time service;
  • · Space tracking and navigation;
  • · Metrology and testing;
  • · Scientific research.

Specifications

Waveform Qty Amplitude Bandwidth, ps Rise time, ns Interface / Load
5 MHz (sine) 2 1 ± 0.2 VRMS - - SMA/50Ω
10 MHz (sine) 2 1 ± 0.2 VRMS - - SMA/50Ω
Performance
Relative frequency accuracy ±3×10⁻¹³ ≤ ±1×10⁻¹³
Frequency stability (Allan deviation at 25±1 °C, environmental effects excluded) 1 s ≤ 3.0×10⁻¹³
10 s ≤ 1.0×10⁻¹³
100 s ≤ 5.0×10⁻¹⁴
3600 s ≤ 9.0×10⁻¹⁵
86400 s ≤ 4.0×10⁻¹⁵
Phase noise
(dBc/Hz)
Offset @5MHz @10MHz @100MHz
1 Hz ≤ -105 ≤ -100 ≤ -80
10 Hz ≤ -130 ≤ -120 ≤ -100
100 Hz ≤ -145 ≤ -140 ≤ -120
1 kHz ≤ -155 ≤ -145 ≤ -150
10 kHz ≤ -155 ≤ -145 ≤ -150
Frequency drift (after 1 year of operation) ≤ 1×10⁻¹⁵/day
Magnetic sensitivity ≤ 1×10⁻¹⁴/Gauss
Temperature sensitivity ≤ 2×10⁻¹⁴/℃
Physical Specifications
Dimensions (W × H × D) 470 mm × 200 mm × 530 mm
Weight ~10kg
Operating temperature range (±0.5 °C/hour) 10–35 °C (recommended +22 °C)
Relative humidity < 80% at +25 °C